Non-Invasive Cardiac Output Assessment by Impedance Cardiography in Patients with Moderate to Severe Aortic valve Stenosis: Comparitive Study with Thermodilution Method
Publication Type
Conference abstract/paper published in a peer review journal
Authors
Fulltext
Download

Aim: The aim of the study was to compare the cardiac output (CO) measured by the impedance cardiography (COicmethod with that of the thermodilution (COth) method in patients with moderate to severe aortic stenosis.
Methods: Measurements of COic and COth were compared in 30 patients with moderate to severe aortic valve stenosis, undergoing diagnostic right and left heart catheterization. Patient age ranged from 37 to 82 years (mean 48); there were 21 men and 9 women. Twenty five patients had sinus rhythm and 5 had atrial fibrillation
For non-invasive measurements of cardiac output by impedance cardiography, standard surface electrodes were applied to the left side of the neck and the left side of the thorax at the level of the xiphoid process. COth was assessed in triplicate by thermodilution via pulmonary artery catheterization
Results: The mean COic was 4,9 ±1,4 litre min−1. The mean COth was 5,0 ±1,6 litre min−1.
There was no significant difference (P=0.6). An well correlation (r=0.73) was found between COic and COth (P<0.001). The bias between the two methods (COic-COth) was 0.03 litre min−1 and the limits of agreement, defined as the bias ±1,96 SD were − 1,9 and +2.0 litre min−1.
Conclusions: Impedance Cardiography (ICG) is a feasible and accurate method for non-invasive measurements of CO. Haemodynamic measurements by ICG were correlated highly significant to measurements by the TD method.

Journal
Title
Clinical Research in Cardiology
Publisher
D. Steinkopff-Verlag
Publisher Country
Germany
Indexing
Thomson Reuters
Impact Factor
5.2
Publication Type
Both (Printed and Online)
Volume
100
Year
2011
Pages
--